Reliable gas measurement starts with understanding the principles behind it. Explore EffecTech’s Knowledge Hub for short explanations of gas quality measurement concepts, methods and common questions. Topics range from measurement basics to calibration, testing and measurement assurance.
Topics
- Metrology basics: accuracy, precision, errors, uncertainty and fitness for purpose
- Measurement units and traceability: SI units and how gas measurement results are linked to recognised references
- Reference gases: CRMs, PRGMs and CGMs, their uses and how to choose a gas mixture that is fit for purpose
- Calibration and validation: how calibration relates instrument response to reference values, and how validation demonstrates fitness for purpose.
Metrology basics
Metrology is the science of measurement. It helps us understand what a measurement result tells us, how reliable it is and whether it is suitable for its intended use.
This section introduces the concepts that underpin reliable gas measurement, including accuracy, precision, measurement errors and uncertainty. These provide a foundation for understanding how measurements are made, checked and compared.
These explanations draw on the definitions in the International Vocabulary of Metrology (VIM), with examples relevant to gas measurement. For the formal definitions and further detail, see the online VIM, hosted by BIPM.
Contents:
- What is metrology?
- What is measurement accuracy and why does it matter?
- Does measurement uncertainty mean a result is unreliable?
- What is precision, and can precise measurements still be wrong?
- What is measurement bias?
- What is the difference between random and systematic measurement errors?
- What is measurement repeatability?
- What is measurement reproducibility?
- Why does measurement uncertainty matter in gas measurement?
- Where does uncertainty in gas measurement come from?
- How are accuracy, precision and uncertainty related?
- What makes a gas measurement system fit for purpose?
What is metrology?
Metrology is the science of measurement. While measurement is fundamental to almost every aspect of our lives, it is probably not something we think about very often. In simple terms, measuring something means assigning a value to a particular property, such as a vehicle’s speed, a person’s weight or a patient’s blood pressure.
Metrology is concerned with how we know that those measured values can be trusted. It provides the principles and methods needed to demonstrate that measurements are sufficiently accurate for their intended purpose, are traceable to recognised standards and have a known level of uncertainty.
Metrology underpins all of EffecTech’s products and services. Whether we are producing calibration gases or assessing gas or LNG sampling systems, we apply metrological principles to ensure gas quality measurements are fit for purpose, have a known uncertainty and are traceable to recognised standards. The following sections explain the key concepts that make this possible.
What is measurement accuracy and why does it matter?
Accuracy describes how close a measured value is to the true value of what is being measured. As the true value is generally not known exactly, measurements are assessed against appropriate reference values.
In everyday situations, small inaccuracies may not matter very much. However, in industrial and commercial gas measurement, accuracy is critical. Even small measurement errors can affect compliance, emissions reporting and commercial transactions.
For example, inaccurate gas quality data could lead to incorrect energy calculations, fuel gas being undervalued or overvalued, or a failure to identify gas falling outside required specifications, such as total sulphur limits for gas entering the transmission system.
EffecTech’s calibration, testing, performance evaluation and inspection services help our customers achieve accurate, reliable and traceable measurements they can trust.
Does measurement uncertainty mean a result is unreliable?
In everyday language, the word “uncertainty” can sound negative or imply doubt. However, in measurement science, uncertainty has a specific technical meaning.
When expressed as an interval, measurement uncertainty describes the range within which the true value is expected to lie, at a stated level of confidence. It helps you judge whether a result is suitable for its intended use.
For example, if a calibration gas certificate states a methane concentration of 90.00 mol% with an expanded uncertainty of ± 0.05 mol%, the true value is expected to lie between 89.95 mol% and 90.05 mol% at the stated level of confidence.
A gas with a lower uncertainty, such as ± 0.01 mol%, provides a narrower range at the same confidence level and is typically used where very high measurement accuracy is required.
A gas with a higher uncertainty, such as ± 0.50 mol%, has a much wider possible value range and may be less suitable for applications where accurate gas quality measurements are critical.
The aim is an uncertainty that is fit for purpose: small enough to meet the application’s requirements. Seeking a lower uncertainty than is needed can add expense without providing a useful benefit.
When uncertainty is properly calculated and reported, it helps users understand the reliability and quality of the measurement.
What is precision, and can precise measurements still be wrong?
Precision describes how closely repeated measurement results agree under specified conditions. For example, if you weigh yourself 10 times on the same bathroom scale and it gives almost exactly the same reading each time, the measurements are precise. If the readings vary significantly, they have poor precision.
However, precision does not guarantee accuracy. If the scale has not been calibrated correctly and always reads 2 kg too high, it introduces a bias into every measurement. The readings may still be very consistent (precise), but they will all be offset from your true weight, making them inaccurate.
The same principle applies to gas measurement. If there is bias in your sampling system or gas chromatograph (GC), the measurements may be very consistent but still consistently wrong. Taking more measurements will not remove that bias. To improve accuracy, the source of the bias must be identified and addressed, for example through a sampling system inspection, calibration with traceable reference gases or an independent performance evaluation.
This is why it is important to understand the difference between precision and accuracy. A measurement system can produce highly consistent results, but if there is bias somewhere in the measurement process, those results may still be wrong.
What is measurement bias?
The International Vocabulary of Metrology defines measurement bias as an estimate of systematic measurement error. Bias can cause measurements to be consistently higher or lower than the true value. It can occur if there is a problem with the measuring system, the sampling system, the instrument, its calibration or set-up, or through the way the measurement is carried out. This systematic effect influences the accuracy of the measurement.
Bias is different from a random error, which varies unpredictably from one measurement to the next. Repeating a measurement many times and averaging the results may reduce the effect of random errors, but it will not remove bias. Instead, the underlying cause must first be identified and addressed.
What is the difference between random and systematic measurement errors?
Measurement errors fall into two categories: random errors and systematic errors.
Random errors vary unpredictably between measurements. They affect the precision (repeatability and reproducibility) of measurements because repeated tests may produce slightly different results, scattered around the mean value. Random variation can be reduced by good technique, but never completely eliminated. Its contribution to measurement uncertainty can be estimated by carrying out repeated tests.
Systematic errors remain constant or vary predictably under given conditions. They can arise from issues somewhere in the measurement process, such as incorrect gas chromatograph (GC) set-up, calibration problems or issues within the sampling system. They can cause results to differ systematically from the true value, even when random errors are very small. Bias is an estimate of this systematic error. Simply repeating the test does not reveal a bias shared by all the results.
If systematic errors are not identified and addressed, they can lead to ongoing inaccuracies that may affect compliance, operational decisions and commercial gas calculations over time. Both random and systematic errors can arise from the way measurements are carried out or from issues with the instrument or sampling system.
EffecTech’s calibration, inspection and performance evaluation services help identify both random and systematic sources of error to improve measurement confidence and reliability.
What is measurement repeatability?
Repeatability is the consistency of results obtained by the same person using the same equipment and procedure, in the same location and under the same operating conditions, over a short period of time.
Repeatability demonstrates the ability of the instrument and the measurement process to produce closely agreeing results consistently. However, good repeatability alone does not demonstrate accuracy.
What is measurement reproducibility?
Reproducibility is the consistency of results obtained by different people using different equipment at different locations, such as different laboratories. It helps show whether closely agreeing results can be obtained across those measurement settings.
Repeatability concerns measurements made under the same conditions over a short period of time; reproducibility concerns measurements made across different settings.
Why does measurement uncertainty matter in gas measurement?
When expressed as an interval, measurement uncertainty describes the range within which the true value is expected to lie, at a stated level of confidence. It helps you judge whether a result is suitable for its intended use.
In gas quality measurement, even small uncertainties can have significant commercial, operational and regulatory impacts. They can affect:
- Energy calculations and billing
- Custody transfer agreements
- Compliance with gas specifications
- Emissions reporting
- Process efficiency and safety
For example, if the uncertainty associated with a gas chromatograph is too large, it may not be possible to demonstrate whether gas meets a required specification or contractual limit.
Understanding and controlling uncertainty helps operators make informed decisions, reduce risk and maintain confidence in their measurement systems.
EffecTech’s calibration, testing and performance evaluation services help customers quantify and control measurement uncertainty, supporting accurate, reliable and traceable gas quality data that is fit for purpose.
Where does uncertainty in gas measurement come from?
Every stage of the measurement process can contribute to the overall measurement uncertainty. This includes the sampling system, the analytical instrument, the calibration gas, environmental conditions and the calibration procedure itself. These contributions need to be evaluated together to understand how well the reported result represents the gas composition.
If any part of the measurement system introduces bias, such as a poorly performing sampling system or an incorrectly calibrated gas chromatograph (GC), the measurement may also become less accurate. Understanding and controlling these sources of uncertainty and bias is essential because the final measurement result is only as reliable as the complete measurement system on which it is based.
How are accuracy, precision and uncertainty related?
Accuracy describes how close a measured value is to the true value, while precision describes how closely repeated measurement results agree under specified conditions. When expressed as an interval, uncertainty describes the range within which the true value is expected to lie, at a stated level of confidence.
These concepts are interrelated. A measurement system may produce highly repeatable results (good precision), but if there is a systematic error, the results may still be inaccurate. Uncertainty helps quantify how well the measured value is known, taking account of random variation and contributions from systematic effects.
In gas measurement, accuracy, precision and uncertainty are important because even small errors can affect compliance, emissions reporting, energy calculations and commercial transactions. EffecTech’s calibration, testing and performance evaluation services help customers achieve accurate, precise and traceable measurements they can trust.
What makes a gas measurement system fit for purpose?
A measurement system is fit for purpose when it has been shown to meet the requirements of its intended application. This means providing the appropriate level of evidence and metrological assurance for the measurement being made, so that the results can be relied upon with confidence without introducing unnecessary complexity or cost. A fit-for-purpose measurement is one that is demonstrably capable of supporting the decisions it is being used to make.
At EffecTech, this means ensuring that gas quality measurements are supported by appropriate calibration, traceability, known measurement uncertainty and independent verification where required. By demonstrating that their gas quality measurement system is fit for purpose, organisations can have confidence that the data they rely on for operational decisions, contractual obligations and compliance is both technically sound and defensible.
The same principle applies when selecting a reference gas. For most instrument calibration applications, a suitable Certified Reference Material (CRM) provides the appropriate level of evidence and metrological assurance because its certified values, uncertainty and traceability have already been established. A Primary Reference Gas Mixture (PRGM) is primarily used as a reference standard to establish traceability and reference values for other gas mixtures, rather than for calibration of analytical instruments. Unless the application specifically requires the lower uncertainty available from a PRGM, a CRM with an appropriate composition and uncertainty is usually the most practical and cost-effective choice.
Back to metrology contents | Back to topics
Measurement units and traceability
SI units provide a common basis for measurement, while traceability connects measurement results to recognised references. These short explanations show how both support reliable gas quality measurements. They draw on the BIPM’s International System of Units (SI) brochure and the International Vocabulary of Metrology (VIM), with gas measurement examples to explain their practical application.
Contents:
- What are SI units and how are they used in gas measurement?
- What is amount fraction in gas measurement?
- What does traceability mean in gas measurement?
- Does traceability guarantee that a measurement is accurate?
- How does a traceable calibration gas support reliable gas measurements?
What are SI units and how are they used in gas measurement?
The International System of Units (SI) was established in 1960. It now provides globally agreed definitions of seven base units: the second (time), metre (length), kilogram (mass), ampere (electric current), kelvin (thermodynamic temperature), mole (amount of substance) and candela (luminous intensity). Together, these provide the foundation from which all other SI units are derived.
In gas measurement, one of the most important base units is the mole, which expresses amount of substance. Gas composition is typically reported as an amount fraction, also called mole fraction, expressed for example in mol% or µmol/mol. This makes the mole fundamental to gas analysis and calibration.
The mole is defined using the Avogadro constant, a fixed value that specifies the exact number of elementary entities (such as atoms or molecules) in one mole. This means that one mole represents the same number of entities everywhere in the world. It provides a common reference point for preparing, certifying and comparing gas mixtures, and forms the basis for establishing traceability to the SI.
Back to units and traceability contents
What is amount fraction in gas measurement?
Amount fraction, also called mole fraction, describes the proportion of a particular component in a gas mixture. It is the amount of that component, measured in moles, divided by the total amount of all components in the mixture.
For example, a methane amount fraction of 90 mol% means that methane accounts for 90 out of every 100 moles of the mixture. Smaller proportions are often expressed in micromoles per mole (µmol/mol): 10 µmol/mol means 10 moles of that component for every million moles of the mixture.
Stating the basis clearly matters because amount fraction, mass fraction and volume fraction describe composition in different ways and should not be treated as interchangeable.
Back to units and traceability contents
What does traceability mean in gas measurement?
Measurement traceability is the property of a measurement result that allows it to be related to a recognised reference through a documented, unbroken chain of calibrations, each contributing to the overall measurement uncertainty. For most scientific and industrial measurements, that recognised reference is the International System of Units (SI).
In gas measurement, primary methods such as gravimetric preparation (which involves accurately weighing the gases added to a mixture) are used to produce Primary Reference Gas Mixtures (PRGMs) with values traceable to the SI. National Metrology Institutes (NMIs) maintain primary gas standards, while the Bureau International des Poids et Mesures (BIPM) coordinates international comparisons to demonstrate agreement between them. PRGMs can then be used as references to establish the certified values of other gas mixtures, including Certified Reference Materials (CRMs) supplied for instrument calibration.
Each stage in the traceability chain is linked to the reference through documented measurements, with the associated uncertainty evaluated and recorded at every step. This provides objective evidence that the certified value assigned to a calibration gas can be related back to internationally recognised standards.
For most users, this traceability is largely invisible. When you purchase a CRM, the traceability of its certified values has already been established and is documented on the certificate. The benefit is confidence that the gas composition has been determined using internationally recognised metrological principles. Together with an uncertainty appropriate to the application, this supports its use as a reliable reference for calibration.
Back to units and traceability contents
Does traceability guarantee that a measurement is accurate?
No. Traceability demonstrates that a measurement result can be related to a recognised reference through a documented, unbroken chain of calibrations. It does not, by itself, guarantee that the result is accurate or that its uncertainty is small enough for the intended use.
For example, a gas chromatograph (GC) may be calibrated using a traceable reference gas, but a problem in the sampling system could change the composition of the gas before it reaches the instrument. The reported result may then fail to represent the gas being sampled.
Reliable gas measurements therefore require both traceability and a measurement system that performs correctly, with uncertainty appropriate to the decisions being made.
Back to units and traceability contents
How does a traceable calibration gas support reliable gas measurements?
The traceability chain does not end with the calibration gas. When a gas chromatograph (GC) is calibrated using a Certified Reference Material (CRM) with documented SI traceability, the measurement results produced by that instrument can also be traceable to the SI. This requires the calibration to be carried out correctly and documented, the measurement uncertainty to be evaluated, and the measurement system to continue performing as intended.
However, calibration is only one part of the measurement process. The final measurement result also depends on the performance of the analytical instrument, the sampling system, the calibration procedure and the way the instrument is operated and maintained. If any of these introduce errors or bias, the measurement result may no longer accurately represent the gas being measured, even though the calibration gas’s certified values remain traceable.
This is why traceability is central to all of EffecTech’s products and services. Traceable calibration gases establish the reference point for measurement, while inspections and performance evaluations help assess whether the measurement system produces results that are fit for purpose under real operating conditions.
Back to units and traceability contents | Back to topics
Reference gases
Reference gases provide known composition values against which gas measurement instruments can be calibrated and checked. This section explains the different types supplied by EffecTech and how their uncertainty, traceability, stability and certification help determine which is appropriate for an application.
These explanations draw on the International Vocabulary of Metrology (VIM), ISO 17034 (the international standard setting out requirements for the competence of reference material producers), and EffecTech’s calibration and certification procedures.
Contents:
- What is a Certified Reference Material (CRM)?
- What is a Primary Reference Gas Mixture (PRGM) and who uses one?
- Why is a CRM usually the right choice for instrument calibration?
- What is a Calibrated Gas Mixture (CGM) and when is it supplied?
What is a Certified Reference Material (CRM)?
A Certified Reference Material (CRM) is a reference material supplied with a certificate stating its certified values, their uncertainty and metrological traceability. In gas measurement, CRMs are widely used to calibrate gas chromatographs and other analytical instruments because they provide an effective balance between low uncertainty, traceability and cost.
CRM production involves internationally recognised procedures for establishing certified values and assessing the mixture’s homogeneity and stability. ISO 17034 sets out the requirements for the competence of reference material producers. For gas CRMs produced under this accreditation, the certificate includes a shelf life or validity period, indicating how long the certified values remain valid under the specified storage and use conditions.
CRMs are widely used in applications where high measurement confidence, demonstrated stability and support for regulatory compliance are required.
EffecTech produces CRMs under ISO 17034 accreditation. Where the gas mixture falls within both its ISO 17034 and ISO/IEC 17025 accreditation scopes, EffecTech supplies both a reference material certificate and a calibration certificate as standard, supporting customers who require evidence of either or both accreditations.
Back to reference gases contents
What is a Primary Reference Gas Mixture (PRGM) and who uses one?
Primary Reference Gas Mixtures (PRGMs) are gas standards whose composition is established using a primary measurement method, such as high-precision gravimetric preparation, and verified analytically. They sit at the top of the gas measurement traceability hierarchy and can provide some of the lowest available uncertainties in gas composition.
Because producing PRGMs requires highly specialised preparation and verification processes, they are typically more expensive than mixtures intended for instrument calibration. They are mainly used by organisations requiring very low measurement uncertainty, such as National Metrology Institutes (NMIs) and ISO/IEC 17025 accredited laboratories establishing reference values for internal gas standards.
For calibration of gas chromatographs (GCs) used in fiscal metering and process applications, a suitable Certified Reference Material (CRM) is often a more practical and cost-effective choice. Calibrated Gas Mixtures (CGMs) may also be appropriate, depending on the application’s requirements for uncertainty, stability and certification.
Back to reference gases contents
Why is a CRM usually the right choice for instrument calibration?
A Primary Reference Gas Mixture (PRGM) sits at the top of the reference gas hierarchy and is primarily used to establish reference values for other gas mixtures. Producing a PRGM requires specialist expertise, carefully controlled preparation methods and rigorous evaluation of measurement uncertainty.
For most instrument calibration applications, the very low uncertainty available from a PRGM is not required. A suitable Certified Reference Material (CRM) provides certified composition values, documented uncertainty, traceability and stability information that meet the application’s requirements, usually at a lower cost. Selecting a CRM rather than a PRGM is therefore a matter of choosing a reference material that is fit for purpose. Paying for a lower uncertainty than the application needs may add expense without a useful improvement in measurement performance.
Back to reference gases contents
What is a Calibrated Gas Mixture (CGM) and when is it supplied?
A Calibrated Gas Mixture (CGM) is a gas mixture whose composition has been determined by calibration against appropriate reference standards. It is supplied with a calibration certificate documenting the measured values, their uncertainty and metrological traceability.
EffecTech supplies a mixture with a CGM certificate only, without an accompanying CRM certificate, when the requested mixture is covered by its ISO/IEC 17025 accreditation scope but not its ISO 17034 scope.
CGMs can be used to calibrate, validate and test gas measurement instruments, including gas chromatographs (GCs), analysers and emissions monitoring systems, provided the mixture’s composition, uncertainty and stability are suitable for the intended application.
Back to reference gases contents | Back to topics
Calibration and validation
Calibration and validation both support reliable gas measurements, but they have different purposes. This section explains their roles and how an EffecTech performance evaluation contributes evidence that an analytical system is fit for purpose.
These explanations draw on the International Vocabulary of Metrology (VIM) and ISO 10723, which describes how to evaluate the performance of natural gas analytical systems.
Contents:
- What is calibration in gas measurement?
- What is validation in gas measurement?
- What is the difference between calibration and validation?
- How does an EffecTech performance evaluation support validation?
What is calibration in gas measurement?
Calibration establishes the relationship between an instrument’s response and the known values of a reference standard, taking account of measurement uncertainty. For gas chromatographs (GCs), this involves analysing a calibration gas with a certified composition and using the instrument’s response to establish how gas composition is calculated.
Calibration information can be used to apply corrections to measurement results or identify whether the instrument needs adjustment. Adjustment is a separate operation and is normally followed by recalibration. These activities support accurate measurements, but calibration alone does not confirm that the instrument meets the requirements of its intended use.
The final measurement result also depends on factors such as the sampling system, instrument performance, maintenance and operating practices. Calibration is therefore one part of the evidence used to demonstrate that a measurement system is fit for purpose.
An EffecTech performance evaluation builds on this calibration information by assessing the GC’s performance across a defined range of gas compositions.
Back to calibration and validation contents
What is validation in gas measurement?
In everyday use, validation is often used to mean “proving that something works”. In gas measurement, it means demonstrating, using objective evidence, that the requirements for an intended application are appropriate and have been met. This provides confidence that the resulting measurements are fit for purpose.
Validation can apply to a measurement method, an instrument or a complete measurement system. When validating a gas measurement system, the evidence may need to cover the sampling system, the analytical instrument, the calibration procedure, operating conditions and other factors that could affect the final result. For example, the measurement uncertainty must be small enough to support the operational, contractual or compliance decisions being made.
Validation can draw on evidence from a range of activities, such as calibration, inspections, performance evaluations and quality control procedures. For example, an EffecTech performance evaluation provides evidence of whether a GC can meet the application’s measurement requirements across a defined range of gas compositions. The evidence needed depends on the application: the aim is to demonstrate that the system meets its intended requirements, rather than simply that individual checks have been completed.
Back to calibration and validation contents
What is the difference between calibration and validation?
Calibration and validation have different objectives. Calibration establishes the relationship between an instrument’s response and known reference values, including their uncertainty. Validation demonstrates that the requirements for an intended application are appropriate and have been met.
For example, calibrating a gas chromatograph (GC) establishes how its response relates to the certified composition of a reference gas. Validating the gas measurement system requires evidence that it can deliver results suitable for its intended application. This may include calibration information, sampling system performance, instrument performance, operating conditions and the overall measurement uncertainty.
Calibration establishes how an instrument’s response relates to known reference values. Validation demonstrates that the measurement system is suitable for its intended purpose.
Back to calibration and validation contents
How does an EffecTech performance evaluation support validation?
An EffecTech performance evaluation provides objective evidence of how a gas chromatograph (GC) performs across a defined range of gas compositions. Using traceable reference gases and the method set out in ISO 10723, it assesses the instrument’s response, repeatability, measurement errors and uncertainty.
The results can be compared with the requirements of the intended application, such as contractual limits or regulatory specifications. This helps establish whether the GC, using its specified calibration gas, can deliver measurements that are fit for purpose across that composition range.
A performance evaluation therefore contributes directly to validation of the analytical system. To demonstrate that the complete gas measurement system is fit for purpose, its findings need to be considered alongside evidence about the sampling system and other relevant operating conditions.
Back to calibration and validation contents | Back to topics